48T
$299.00
96T
$399.00
Species:
Mouse
Alternative Name:
Lysosomal acid lipase/cholesteryl ester hydrolase, Acid cholesteryl ester hydrolase, LAL, Cholesteryl esterase, Lipase A, Sterol esterase, Lipa, Lip1
Uniport ID:
Q9Z0M5
Assay Type:
Sandwich
unit:
ng/mL
Standard:
10ng/mL
Sensitivity:
0.1 ng/mL
Detection Range:
0.16-10ng/mL
Assay Time:
3.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Research:
CD & Adhesion molecule;
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Mouse LIPA. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse LIPA and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse LIPA, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Mouse LIPA.You can calculate the concentration of Mouse LIPA in the samples by comparing the OD of the samples to the standard curve.
| ng/mL | OD | Corrected |
|---|---|---|
| 10 | 2.382 | 2.358 |
| 5 | 1.697 | 1.673 |
| 2.5 | 0.981 | 0.957 |
| 1.25 | 0.474 | 0.45 |
| 0.63 | 0.235 | 0.211 |
| 0.32 | 0.112 | 0.088 |
| 0.16 | 0.065 | 0.041 |
| 0 | 0.024 | 0 |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean(ng/mL) | 0.2 | 1.79 | 2.69 | 0.18 | 2.41 | 2.93 |
| Standard deviation | 0.01 | 0.07 | 0.21 | 0.01 | 0.09 | 0.22 |
| C V (%) | 6.46 | 3.87 | 4.12 | 7.07 | 6.1 | 6.2 |
| Sample Type | Range(%) | Average Recovery(%) |
|---|---|---|
| Serum (n=8) | 80-95 | 87 |
| EDTA plasma (n=8) | 83-95 | 89 |
| Cell culture media (n=8) | 90-105 | 97 |
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 86-95 | 91-99 | 85-96 |
| Average (%) | 90 | 98 | 86 | |
| 1:4 | Range (%) | 93-103 | 83-97 | 87-101 |
| Average (%) | 98 | 89 | 89 | |
| 1:8 | Range (%) | 85-94 | 87-105 | 89-102 |
| Average (%) | 86 | 99 | 95 | |
| 1:16 | Range (%) | 87-101 | 87-99 | 85-98 |
| Average (%) | 99 | 90 | 91 |